Oxygen analyzer

Gasmet 19” frame analyzer with microprocessor technology has the ZrO2 cell inside and can be used for O2 measurement in CEM II or EEMS cabinets or as a portable analyzer.

Together with a sampling and conditioning system Gasmet oxygen analyzer can measure the O2 content in wet flue gas or steam down to less than 1 ppm O2. Gasmet oxygen analyzer is manufactured by ENOTEC.

Technical Data

General parameters
Measuring principle: ZrO2 measuring cell
Detection limit: < 1 ppm O2
Response time: < 1 second
Sample gas temperature: 120 °C to 300 °C, non-condensing
Environmental temperature: -20 °C to +40 °C
Line voltage: 115 or 230 V / 50 -60 Hz
Reference gas: Instrument air, dew point less than –40 °C, no oil, about 30 l/h
Measuring gas: Dry or wet, no combustibles
Calibration gas: Instrument air as above or test gas from bottles

 

 ZrO2 sensor and SME5 electronic board
Sensor material: 316 ss (1.4571)
Gas connection: 6 mm Swagelok
Gas flow: Minimum 0.1 l/min
Power consumption: 150 VA during heating, 60-100 VA continuously
Heater voltage: 48 VAC for ZrO2 sensor
Fuse requirement: 10 A
Measurement ranges: Two (2), key selectable, each from 0-2 % O2 to 0-25 % O2 programmable, other on request
Accuracy: +/-0.2 % of reading, maximum
Operation: With softkeys, menu-controlled self-explanatory
Display: LCD display LED illuminated 240 x 64 dots
Spectrum: Actual values, parameters, system functions, system status
Language: German and English
Alarm display: LED for alarm, fault and maintenance
O2 value memory: On/off by program during calibration the last O2 value kept in memory
System tests: Thermocouple polarity
Data fault EEPROM
Electronic temperature
ZrO2 cell temperature
ZrO2 cell mV-value
Automatic start up with single point Calibration with air
Calibration: Single point with air, two point with air and test gas bottle
Analogue outputs: O2 measured value
0/4-20 mA, active sourcing, DC-isolated
Digital output: Error status, maintenance, measuring range, measuring limit 1, measuring limit 2 (relay contact)

 

Enclosure
Material: Aluminium
Dimensions (mm): 177 * 483 * 400
Weight: 12 kg
TÜV approval: TÜV approved in combination with a multicomponent analyser Gasmet CX4000
EMC tested: In accordance with EN50082-2, EN55011 Cl.B



Gasmet Technologies Oy shall not be liable for technical or editorial errors or omissions contained herein. The information in this document is provided “as is” without warranty of any kind and is subject to change without notice. Should you find any errors, we would appreciate if you notified us.

Videos

This Product does not have any videos.