Gasmet In-Situ

Gasmet™ In-Situ analyzer consists of a sample cell inserted into the stack or duct, heated steel mounting flange and the rugged GICCOR™ interferometer unit in an IP65 enclosure that is directly attached to the flange.

Gasmet™ In-Situ offers a cost-effective alternative for extractive systems in continuous emissions & process monitoring, where multiple compounds are monitored. Typically, concentrations of H2O, CO2, CO, NO, NO2, N2O, SO2, HCl, HF, NHand CH4 are continuously monitored with a single Gasmet™ In-Situ. In addition, different VOC’s can be measured as individual compounds. There is no need for any span calibrations; only zero calibration with instrument air every 24 hours. Measurement results can be transmitted with analog 4–20 mA signals or with a digital protocol (MODBUS, PROFIBUS etc.).

By combining powerful FTIR technology with rugged in-stack measurement cell, the Gasmet™ In-Situ offers increased analytical capability and reduced cross-interference effects when compared with other types of in-situ analyzers.

 

SYSTEM PARTS

  • GasmetTM In-Situ
  • Auxiliary unit with I/O functions and pneumatics included
  • GasmetTM industrial computer with RS-422 connection
  • Instrument air
  • Vortex cooling as an option
  • Additional I/Os

KEY ADVANTAGES

  • Non-extractive technique
  • Advantages of FTIR technology
  • Very resistant to stack movement and vibrations

AREAS OF APPLICATIONS

  • Aluminum smelters

Technical Data

General parameters
Measuring principle: Fourier transform infrared, FTIR
Performance: Simultaneous analysis of up to 50 gas compounds
Response time, T90: Typically < 360 s, depending on the gas flow and measurement time
Operating temperature: -30 – +40 °C, non condensing
Storage temperature: -30 – +60 °C, non condensing
Power supply: 100-115 or 230 V / 50 -60 Hz
Power consumption: 2200 W max., 500 W continuous (with cell heating)
Instrument air: 4-7 bar pressure, dew point –20 oC, oil and dust free
Max. air consumption: 250 l/min for vortex cooling (cont.)
100 l/min for zero calibration/purge
(15 minutes at 24 hour intervals)

 

Spectrometer
Resolution: 8 cm-1
Scan frequency: 10 scans / s
Detector: Peltier cooled MCT
Source: SiC, 1550 K
Beamsplitter: ZnSe
Window material: ZnSe
Wave number range: 900 - 4 200 cm-1

 

Sample cell
Structure: Multi-pass, fixed path length 6.0 m
Probe body material: AISI 316 steel
Mirror material: Rhodium coated aluminium
Mirror coating: Reflective gold coating
Gaskets: Viton® O-rings
Window material: BaF2
Insertion depth: 736 mm
Mounting flange: 240 mm diameter
Temperature: 250 °C
Gas filtration: Filter panels (2-7 μm)

 

Measuring parameters
Zero point calibration: 24 hours, calibration with instrument air
Zero point drift: < 2 % of measuring range per zero point calibration interval
Sensitivity drift: None
Linearity deviation: < 2 % of measuring range
Temperature drifts: < 2 % of measuring range per 10 K temperature change
Pressure influence: 1 % change of measuring value for 1 % sample pressure change. Sample pressure changes are measured and compensated

  

 Sample gas conditions
Gas temperature: Up to 250 °C, non-condensing

 

Electrical connectors & electronics
Digital interface: RS-232/422 cable for the analyser.
Power connection: Terminal connectors for L + N + PE wires
A/D converter: Dynamic range 95 dB
Signal processor: 32-bit floating point DSPs, 120 MFLOPS

 

 Computer unit
Analyzer is connected to an external computer via serial cable. The external computer controls the Gasmet In-Situ analyzer. For I/O options, see Gasmet Industrial Computer Data Sheet.
Minimum configuration: 512 MB Memory, > 40 GB hard disk, > 2.4 GHz Intel Pentium IV processor, modem, network card
Operating system: Windows 7 (32-bit)
Software included: Calcmet for Windows
Watchdog support: Included
Mouse: Included
Keyboard: Not Included
Display: TFT display
Digital Interface: RS-232 cable to analyzer (<5m)
RS-422 option up to 1000m
Remote control: Built-in modem/network card and PC Anywhere support

 

 Options
Adapters for flanges larger than 240 mm
RS-422/485 interface instead of RS-232 for longer distance, maximum distance one (1) kilometer.

 

Enclosure
Material: Aluminium / steel / polyethylethylketone (PEEK)
Dimensions (mm): 1168 x 417 x 263 mm
Weight: 30 kg
CE label: According to EMI guideline 89/336/EC

 